DEPUY SYNTHES POWER TOOLS AIR OSCILLATOR; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 511.610
Device Problems Air Leak (1008); Device Stops Intermittently (1599); Failure to Align (2522); Device Operates Differently Than Expected (2913); Mechanical Jam (2983); Power Problem (3010); Appropriate Term/Code Not Available (3191); Noise, Audible (3273)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/16/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).The device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.The assignable root cause was determined to be due to improper maintenance, which is user error/misuse/abuse.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Event Description
It was reported by (b)(6) that during service and evaluation, it was observed that the air oscillator device motor had seized, jammed, and was heavy moving.It was also noted that the device failed pre-repair diagnostic tests for immediate stopping, symmetry, frequency, starting behavior, performance, excessive noise, and air leak.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIR OSCILLATOR
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ   CH4437
Manufacturer Contact
mark vornheder
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key6140650
MDR Text Key61607876
Report Number8030965-2016-15747
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Service and Testing Personnel
Type of Report Initial
Report Date 09/16/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/02/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number511.610
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/11/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received11/11/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/17/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-