Model Number FUSION |
Device Problem
Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/14/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient information was not made available from the site.On 11/16/2016 a medtronic representative performed a navigation system check-out, all areas passed.System performed as intended.- no parts have been received by manufacturer for analysis.- no further issues have been reported.
|
|
Event Description
|
A medtronic ent representative reported that, while in an ear, nose & throat (ent) procedure, the site's navigation system became unresponsive unexpectedly.This issue occurred mid-procedure.The navigation system was re-booted and normal function was restored.The surgeon opted to continue and completed the procedure with the use of the navigation system.Delay in therapy was less than one hour.There was no impact on patient outcome.
|
|
Manufacturer Narrative
|
A review of the software logs were completed.This anomaly is tracked through test track (b)(4) and references to this case have been added to that record.
|
|
Manufacturer Narrative
|
Product, unique device identification (udi) and device manufacture date updated to proper value.
|
|