DEPUY SYNTHES POWER TOOLS OSCILL-SAW; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 518.010 |
Device Problems
Bent (1059); Break (1069); Device Operates Differently Than Expected (2913); Failure to Shut Off (2939)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.The assignable root cause was determined to be due to improper handling, which is user error/misuse/abuse.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Event Description
|
It was reported by slovenia that the oscillating saw device works all the time, it does not stop.During service and evaluation, it was noted that the handle and lever were deformed and bent, and the valve lifter and handle were bent, the handpiece was out of lubrication, so the motor vanes were damaged.It was also noted that the device failed pre-repair diagnostic tests for operating assessment of the trigger/lever.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available for use.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|