• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number S7
Device Problems Device Displays Incorrect Message (2591); Communication or Transmission Problem (2896); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/10/2016
Event Type  malfunction  
Manufacturer Narrative
Return requested.Replacement axiem portable shipped to site 11/10/2016.No parts have been received by manufacturer for analysis.On 11/11/2016 a medtronic representative performed a navigation system check-out, software and instruments areas passed.Hardware test failed.Axiem communicator not picking up instruments in the field despite having "green" status, replaced axiem communicator.Issue resolved.System performed as intended.
 
Event Description
A medtronic representative reported that, while in a cranial shunt procedure, the surgeon experienced difficulty tracking the non-invasive patient tracker.They successfully registered the patient and tracked the registration probe, however, when they went to navigate the non-invasive patient tracker (nipt) went to red status.The axiem box and emitter were both showing in green status in the tracking details.Trouble-shooting did not resolve the issue.An alternate navigation system was brought in to be used to continue the procedure to completion.Delay in therapy was less than one hour.There was no impact on patient outcome.
 
Manufacturer Narrative
The axiem box was returned to the manufacturer for analysis.The device tracked properly with no failures.Visual inspection did find physical damage to the usb connector connecting to the computer, however this did not affect function.The device was found to be fully functional.The reported event could not be duplicated by medtronic personnel.The separate issue found, regarding physical damage, was documented in a medtronic navigation hardware anomaly tracking database.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
amos jarrette
826 coal creek circle
louisville, CO 80027-9710
7208902082
MDR Report Key6150024
MDR Text Key61615989
Report Number1723170-2016-05668
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450944
UDI-Public00613994450944
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 01/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/06/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberS7
Device Catalogue Number9733856
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received12/21/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/09/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age59 YR
Patient Weight75
-
-