A steris service technician arrived onsite to inspect the sterilizer.During the technician's inspection, he found that a contactor on the generator element was not operating properly.The technician identified that the contactor was stuck in a closed position causing the integral steam generator on the unit to continue running.As the steam generator continued to run, steam and water was released from the unit's safety valve.The steris service technician replaced the contactor, steam generator, and safety valve; ran a test cycle; and confirmed the unit to be operating properly.The sterilizer was returned to service.
|