Model Number MT22719-PNK
Device Problem No Audible Alarm (1019)
Patient Problem Injury (2348)
Event Date 09/08/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on (b)(6) 2016, that on (b)(6) 2016, the patient experienced no audio output from the receiver and an adverse event.It was indicated that medical intervention was required but no details were provided.Additionally, the patient tested the receiver and it did not beep.No further event or patient information is available.The receiver was returned for evaluation.A follow-up will be submitted upon completion of device investigation.
 
Manufacturer Narrative
(b)(4).
 
Event Description
The complaint device was returned for evaluation.The device was visually inspected and no defect was found.Functional testing was performed and the reported fault could not be reproduced and there was no failure detected.A review of the downloaded receiver log did not find any errors related to the customer complaint.The device was determined to be operating within the required specifications without malfunction.The reported event of no audio output was not confirmed.A root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key6153564
MDR Text Key61695068
Report Number3004753838-2016-88450
Device Sequence Number1
Product Code MDS
UDI-Device Identifier00386270000262
UDI-Public(01)00386270000262(241)MT22719-PNK(10)5215855(17)NA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date 11/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/07/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMT22719-PNK
Device Catalogue NumberSTK-GF-PNK
Device Lot Number5215855
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received12/07/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number3004753838-02/29/16-001C
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age33 YR
-
-