A medtronic representative reported that during a spinal fusion procedure, an inaccuracy was alleged against the navigation system.It was reported that at the beginning of the procedure, a 3d scan was acquired using the imaging system.Next, the surgeon verified their instruments and calibrated the instrument tracker to be used with a third party driver and screws.Accuracy was then verified, although an 'abnormality' was noticed.The specifics of this 'abnormality' could not be described by the surgeon.The surgeon made the decision to proceed with the case as planned, and he began placing the first screw.At this point, the surgeon did not feel the screw was being inserted in the correct position, and upon another accuracy check it was discovered that the screw had in fact come out the side of the pedicle.A 1 centimeter inaccuracy was found, in the inferior and lateral (right) directions.After this discovery, a second 3d spin was acquired and accuracy was restored.The procedure was then completed as planned, with no further incident.The alleged malfunction delayed the procedure by less than one hour.No additional details were provided.
|
A medtronic representative went to the site to test the equipment and found the imaging system trackers required re-calibration.After re-calibrating the trackers the hardware, software, and instruments passed the system checkout.The system was found to be fully functional.The software investigation found that the symptom was resolved when imaging system trackers were calibrated.Software is functioning as designed.
|