Model Number EP003994S
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/23/2016
Event Type  malfunction  
Manufacturer Narrative
Product event summary: the lead accessories were returned and analyzed.Product analysis revealed the needle had bends throughout.Suspect needle bent while being advanced through the sheath.Damaged during use.
 
Event Description
It was reported that following transseptal puncture, the needle and sheath were removed and flushed.A piece of lining from the sheath was flushed out.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BROCKENBROUGH NEEDLE
Type of Device
TROCAR
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6158286
MDR Text Key61892232
Report Number9612164-2016-01244
Device Sequence Number1
Product Code DRC
Combination Product (y/n)N
Reporter Country CodeEI
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 09/23/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/09/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/09/2018
Device Model NumberEP003994S
Device Catalogue NumberEP003994S
Device Lot Number211235149
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/30/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/23/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/10/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-