• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number S7
Device Problem Imprecision (1307)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/21/2016
Event Type  malfunction  
Manufacturer Narrative
On 11/25/2016 a medtronic representative performed a navigation system check-out, all areas passed.System performed as intended.- no parts have been received by manufacturer for analysis.- no further issues have been reported.
 
Event Description
A medtronic representative reported that, while in a spine procedure, the surgeon alleged an inaccuracy occurred.They had spun with the imaging system and had placed three screws when the surgeon deemed being 2 millimeters deep on the fourth screw.Before performing another spin they checked the stability of the frame and when they put the probe on it, the frame moved significantly.They tightened down the clamp, and re-spun, the surgeon completed the procedure with the use of the navigation system.Delay in therapy was 45 minutes.There was no impact on patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
judith ericson
826 coal creek circle
louisville, CO 80027-9710
7208902187
MDR Report Key6179198
MDR Text Key62541504
Report Number1723170-2016-05774
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450944
UDI-Public00613994450944
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 12/15/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberS7
Device Catalogue Number9733856
Is the Reporter a Health Professional? No
Date Manufacturer Received11/21/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age54 YR
Patient Weight79
-
-