• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SPINE BLOCKER; INSTINCT JAVA SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SPINE BLOCKER; INSTINCT JAVA SYSTEM Back to Search Results
Model Number N/A
Device Problem Material Deformation (2976)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/15/2016
Event Type  malfunction  
Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.This is report one of two for this event.Reference report 3003853072-2016-00158.
 
Event Description
It was reported that the threads were damaged on two blockers during final tightening.They blockers were removed and replaced.There were no reports of patient injury associated with this event.
 
Manufacturer Narrative
Udi number: (b)(4).The returned blocker was evaluated.The threads were confirmed to be deformed and the cause was determined to be cross-threading due to misalignment of the blocker with the mating screw.A review of the manufacturing records did not identify any issues which would have contributed to this event.The labeling was reviewed and found to contain instructions regarding proper device usage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLOCKER
Type of Device
INSTINCT JAVA SYSTEM
Manufacturer (Section D)
ZIMMER SPINE
23 parvis des chartrons
cite mondiale
bordeaux, cedex 33080
FR  33080
Manufacturer (Section G)
ZIMMER SPINE
23 parvis des chartrons
cite mondiale
bordeaux, cedex 33080
FR   33080
Manufacturer Contact
teresa george
310 interlocken pkwy ste. 120
broomfield 80021
3034437500
MDR Report Key6180760
MDR Text Key62619266
Report Number3003853072-2016-00159
Device Sequence Number1
Product Code NKB
Combination Product (y/n)N
Reporter Country CodeRS
PMA/PMN Number
PK111301
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/16/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number046W0AN00002
Device Lot NumberE86280
Other Device ID Number(01)00889024339910
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received06/25/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/04/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-