AV-TEMECULA-CT NC TREK CORONARY DILATATION CATHETER Back to Search Results
Catalog Number 1012449-15
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 11/21/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Date of event - estimate.Evaluation summary: visual inspection was performed on the returned device.The reported break was confirmed.The investigation was unable to determine a conclusive cause for the reported difficulty.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.Based on the information reviewed, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event Description
It was reported that before use, the nc trek rx 3 x 15 mm balloon dilatation catheter was broken / damaged and not used.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Type of Device
CORONARY DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
EL COYOL, COSTA RICA REG# 3009031392
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key6182921
MDR Text Key62958898
Report Number2024168-2016-08974
Device Sequence Number1
Product Code LOX
UDI-Device Identifier08717648151873
UDI-Public(01)08717648151873(17)190630(10)60727G1
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K110134
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 12/16/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/16/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2019
Device Catalogue Number1012449-15
Device Lot Number60727G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/02/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/22/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-