Patient information was unavailable from the site.A site representative declined to provide patient information.A medtronic representative inspected the navigation system on-site.It was discovered that during the procedure, the surgeon verified one blunt tip pointer prior to use and then used a different blunt tip pointer for navigation (without re-verifying).Also, the site reported that they use a mixture of medtronic spheres along with spheres from other sources to use for navigating instruments.Neither of these particular workflow specifics are recommended by medtronic and both may have caused or contributed to the reported inaccuracy.No parts were replaced or returned to the manufacturer for evaluation.A full navigation system check-out was completed and all tests passed.Full system functionality was confirmed and the system was returned to service.
|
A site representative reported that during a spinal fusion procedure, the navigation system was found to be 10-12 millimeters inaccurate during the accuracy check.It was reported that the surgeon checked accuracy by touching bony anatomy with a blunt tip pointer, and the navigation system displayed the instrument in the spine channel.The surgeon opted to complete the procedure without the use of the navigation system because of this issue.The procedure was completed successfully, and no adverse effect on the patient was reported.
|