Catalog Number RF400J
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Pulmonary Embolism (1498); Death (1802)
Event Date 04/13/2012
Event Type  Death  
Manufacturer Narrative
No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The device was not returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.
 
Event Description
It was reported that approximately two years and 11 months post jugular deployment of a vena cava filter, the patient expired from a pe.It was further reported that the filter was a prophylactic placement to prevent dvt prior to a hip replacement procedure.
 
Manufacturer Narrative
Manufacturing review: the device history records were reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only event reported to date for this lot number and failure mode.Visual inspection and functional/performance evaluation: the device was not returned; therefore, no visual inspection or functional testing of the device was performed.Medical records review: medical records were not provided for review.Image/photo review: images/photos were not provided for review.Conclusion: the device was not returned.Images and medical records were not provided.There was no specific deficiency alleged.The investigation is inconclusive as the origin of the pe was unknown.Based upon the available information, the definitive root cause was unknown.Labeling review: the current ifu (instructions for use) states: potential complications: procedures requiring percutaneous interventional techniques should not be attempted by physicians unfamiliar with the possible complications.Complications may occur at any time during or after the procedure.Possible complications include, but are not limited to, the following: acute or recurrent pulmonary embolism.This has been reported despite filter usage.It is not known if thrombi passed through the filter, or originated from superior or collateral vessels.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that approximately two years and 11 months post jugular deployment of a vena cava filter, the patient expired from a pe.It was further reported that the filter was a prophylactic placement to prevent dvt prior to a hip replacement procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G2 X FILTER SYSTEM - JUGULAR
Type of Device
VENA CAVA FILTER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury NY 12804
Manufacturer Contact
judith ludwig
1625 w 3rd st.
tempe, AZ 85281
4803032689
MDR Report Key6187489
MDR Text Key62784714
Report Number2020394-2016-01179
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082305
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/30/2012
Device Catalogue NumberRF400J
Device Lot NumberGFTC4556
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received01/12/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/17/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age51 YR
Patient Weight136
-
-