• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES, PR SWAN-GANZ BIPOLAR PACING CATHETER FOR SVC INSERTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES, PR SWAN-GANZ BIPOLAR PACING CATHETER FOR SVC INSERTION Back to Search Results
Model Number PE074F5
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/25/2016
Event Type  malfunction  
Manufacturer Narrative
The product is expected to be returned for analysis; however, it has not yet been received.Upon the return of the product a supplemental report will be sent with the investigation results.A device history record review was completed and documented that device met all specifications upon distribution.
 
Event Description
It was reported that during the insertion of a catheter via the internal jugular vein, the catheter "got stuck" in the middle of the procedure.There was no resistance felt against the introducer or the guidewire.Another catheter was inserted from the femoral area and the procedure was finished.There were no patient complications reported.Patient demographic information requested but unavailable.
 
Manufacturer Narrative
One bipolar pacing catheter with attached monoject 1.3 cc limited volume syringe was returned for evaluation.Clotted blood was observed inside the balloon.During evaluation it was able to insert the catheter into a lab 5 fr introflex introducer without issue.The balloon did not inflate due to leakage from a gap around the circumference between the proximal electrode and the catheter body.The leakage was due to an adhesive separation and leakage was from under the proximal electrode.No visible damage to the balloon, windings or returned syringe was observed.Balloon inflation test was performed using returned syringe with 1.3 cc air by holding the balloon under water.Visual examinations were performed under microscope at 20x magnification and with the unaided eyes.Customer report of insertion issue could not be confirmed during the analysis.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint and implement any necessary corrective actions.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised, before deciding to insert or use the catheter, to consider the potential benefits in relation to the possible complications.The techniques for insertion, methods of using the catheter to obtain patient data information, and the occurrence of complications is well described in the literature.No description of the patient¿s anatomy was provided; therefore, it cannot be determined if the patient¿s anatomy, procedural factors or device handling may have contributed to the reported event.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ BIPOLAR PACING CATHETER FOR SVC INSERTION
Type of Device
PACING CATHETER
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
lynn selawski
one edwards way
irvine, CA 92614
9497564386
MDR Report Key6187745
MDR Text Key62794568
Report Number2015691-2016-03815
Device Sequence Number1
Product Code LDF
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K813521
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/09/2018
Device Model NumberPE074F5
Device Lot Number60402658
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/26/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received01/04/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/17/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-