The customer received questionable benz benzodiazepines results for one patient from cobas integra 400 plus serial number (b)(4).The urine result was negative and was reported outside the laboratory.The customer used a value of 200 for the cutoff point for the assay.As the patient was on lorazepam and the result was expected to be positive, the physician asked for confirmation testing.The customer sent the sample for confirmation testing by gc/ms and the result was 1074 ng/ml (positive).The result by gc/ms was believed to be correct.On (b)(6) 2016, the customer repeated the same sample on the cobas integra 400 plus and the result was again negative.The patient was not adversely affected.The customer refused a service visit to check the analyzer.
|
Based on the information provided, a specific root cause could not be determined.As the sample in question was not available, further investigation was not possible.Additional information for further investigation was requested but was not provided.It was suspected a substance in the sample was interfering with the assay.Calibration and qc data were acceptable and did not indicate any reagent issue and confirmed the assay was performing within specifications.
|