• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SPINE PEDICLE SCREW; INSTINCT JAVA SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SPINE PEDICLE SCREW; INSTINCT JAVA SYSTEM Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem Pain (1994)
Event Date 12/01/2016
Event Type  Injury  
Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.The lot number is unknown; therefore the device history records are unable to be reviewed.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.This is report one of two for this event.Reference report 3003853072-2016-00163.
 
Event Description
It was reported that two pedicle screws at l1 were broken in the threaded shaft region outside of the vertebral bodies approximately 14 months post-op.A revision surgery is planned for (b)(6) 2017.
 
Manufacturer Narrative
The device was not returned for evaluation.A photo of the x-ray was provided and shows the screw shaft fractured in the threaded region.The labeling was reviewed and found to contain post-operative warnings and instructions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PEDICLE SCREW
Type of Device
INSTINCT JAVA SYSTEM
Manufacturer (Section D)
ZIMMER SPINE
23 parvis des chartrons
cite mondiale
bordeaux, cedex 33080
FR  33080
Manufacturer (Section G)
ZIMMER SPINE
23 parvis des chartrons
cite mondiale
bordeaux, cedex 33080
FR   33080
Manufacturer Contact
teresa george
310 interlocken pkwy ste. 120
broomfield 80021
3034437500
MDR Report Key6188645
MDR Text Key62839705
Report Number3003853072-2016-00161
Device Sequence Number1
Product Code NKB
Combination Product (y/n)N
Reporter Country CodeRS
PMA/PMN Number
PK111301
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberUNKNOWN INSTINCT PEDICLE SCREW
Device Lot NumberUNKNOWN
Other Device ID NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/25/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age27 YR
-
-