MAUDE Adverse Event Report: SORIN GROUP DEUTSCHLAND S5 GAS BLENDER SYSTEM GAS CONTROL UNIT, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP DEUTSCHLAND S5 GAS BLENDER SYSTEM GAS CONTROL UNIT, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 25-28-67
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 11/24/2016
Event Type  No Answer Provided  
Manufacturer Narrative
Livanova (b)(4) manufactures the s5 gas blender system. The incident occurred in (b)(6). This medwatch report is being filed on behalf of livanova (b)(4). The device was returned to livanova for further investigation. The described error could be reproduced. It was discovered that there was a loose plug connection at the co2 valve. The parts were replaced. The unit was disinfected and cleaned. Preventative maintenance was performed and the unit function properly without any further issue.
 
Manufacturer Narrative
There was no patient involvement. Sorin group (b)(4) manufactures the s5 gas blender system. The incident occurred in (b)(6). This medwatch report is being filed on behalf of sorin group (b)(4). A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue. The gas blender has been returned to sorin group (b)(4) for investigation. A follow-up will be sent when the investigation is complete.
 
Event Description
Sorin group (b)(4) received a report that the s5 gas blender system displayed an error code during installation. There was no patient injury reported.
 
Manufacturer Narrative
The model number provided in the initial report (25-40-45) was incorrect. The correct model number is 25-28-67.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameS5 GAS BLENDER SYSTEM
Type of DeviceGAS CONTROL UNIT, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich,
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer Contact
joan ceasar
14401 w 65th way
arvada, CO 80004
2812287260
MDR Report Key6191920
MDR Text Key250173565
Report Number9611109-2016-00888
Device Sequence Number1
Product Code DTX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K101046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date 10/11/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/20/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number25-28-67
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/01/2016
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/28/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/07/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-