The device was returned to stryker sustainability solutions for evaluation.Visual inspection of the returned complaint device revealed signs of clinical use.The inner and outer shafts of the device were easily separated from each other.There were some signs of galling to the distal end of the outer shaft.Several teeth have been dulled.The inner shaft tip was broken off and cutting tooth damage with galling on the outer surface was evident.The inner shaft tip had several damaged teeth visible.Some metal particulates were visible on outer and inner shaft swabs.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.Therefore, the most likely root causes are the user applied too much force to the device (side-loaded) or the arthroscopic shaver may have come into contact with staples, clips or another metal object, resulting in damage to the blade.The instructions for use (ifu) state: before beginning the procedure, verify compatibility of all instruments and accessories.Plug in and set up the generator according to the instructions in the manufacturer¿s manual.Select an arthroscopic shaver with size, blade, and function most appropriate for the procedure.Inspect the instrument for overall condition and physical integrity.Do not use the instrument if any damage is noted.Return the instrument and packaging to stryker sustainability solutions if it is not in acceptable condition for surgery.Careful handling of the instrument is necessary to avoid damage or breakage as a result of excessive force.Do not apply excessive pressure or ¿side-load¿ the blade during use.Side-loading does not improve the performance of the instrument, can dull the blade, and/or produce metal particulates.Do not allow the arthroscopic shaver to come into contact with staples, clips or any metal object to avoid damage to the blade and possible patient injury.The tip of the bur or cutter must be irrigated periodically (general recommendation: once a minute) to cool the blade and prevent excised tissues from accumulating.Do not run the instrument without appropriate suction for the duration of the process.The reported event will continue to be monitored through post-market surveillance.
|
It was reported during a knee arthroscopy, a piece of the cutter broke inside the patient.The physician was able to retrieve the piece with biopsy forceps and grasper, leading to a 30-45 minute delay.There was no patient injury or medical intervention reported.These are commonly used devices that are readily available.
|