Catalog Number 1013064-06
Device Problem Material Rupture (1546)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/07/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4). During processing of this complaint, attempts were made to obtain complete event, patient and device information. The traveler rx is currently not commercially available in the u. S. However, it is similar to a device sold in the us. The device was not returned for evaluation. A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event. Additionally, a review of the complaint history identified no other incidents from this lot. Based on the information reviewed, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device. The investigation determined the reported difficulty appear to be related to circumstances of the procedure.
 
Event Description
It was reported that the procedure was to treat a lesion located in the mid right coronary artery (rca) with moderate tortuosity and heavy calcification that was 99% stenosed. Resistance was not felt during advancement of the 1. 50 x 6 mm traveler rx balloon dilatation catheter (bdc) and it crossed the lesion successfully. The balloon ruptured during the second inflation at 12 atmospheres. The procedure was completed successfully after a stent was implanted. There were no adverse patient effects or clinically significant delay in the procedure reported. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameTRAVELER RX CORONARY DILATATION CATHETER
Type of DeviceCORONARY DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
EL COYOL, COSTA RICA REG# 3009031392
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key6194315
MDR Text Key63321414
Report Number2024168-2016-09103
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K103110
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation
Remedial Action Other
Type of Report Initial
Report Date 12/21/2016
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/21/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date08/31/2018
Device Catalogue Number1013064-06
Device Lot Number50918G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/12/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-