Catalog Number 6007-008-000 |
Device Problem
Disassembly (1168)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 12/06/2016 |
Event Type
malfunction
|
Event Description
|
It was reported that during testing conducted at the user facility the battery housing of the device was found to be broken off.No patient involvement or procedural delays were reported with this event.
|
|
Manufacturer Narrative
|
The reported event that the battery housing is broken was broken off was confirmed through visual inspection.Any physical impact to the navigated instrument can cause product damage or operational failure due to battery movement.
|
|
Event Description
|
It was reported that during testing conducted at the user facility the battery housing of the device was found to be broken off.No patient involvement or procedural delays were reported with this event.
|
|