Catalog Number SJ-05501
Device Problem Difficult to Remove (1528)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/12/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has not been returned for investigation at this time.The manufacturer will continue to monitor and trend related events.
 
Event Description
The anesthesiologist reported that the epidural catheter has formed a true knot in the tip of the catheter prior to removal, making it extremely difficult to remove.The patient's condition was reported as fine.
 
Manufacturer Narrative
(b)(4).No lot number was provided.A device history record review was performed based upon a lot number from sales history data.A device history record review was performed on the epidural catheter with no relevant findings.The ifu for this kit, e-17019-100d; rev.02, was reviewed as a part of this complaint investigation.The ifu warns the end user, "never advance the catheter more than 5 cm.Advancing the catheter more than 5 cm increases the likelihood of the catheter tip being placed into the anterolateral portion of the epidural space and increases the potential for the catheter knotting." the ifu warns the user, "never tug or quickly pull on catheter during removal from patient to reduce risk of catheter breakage.Do not apply additional tension on the catheter if catheter begins to stretch excessively.Reposition patient to open the vertebral interspaces and re-attempt removal if resistance is encountered or if catheter stretches excessively during removal.During epidural catheter removal, the literature indicates a force of approximately 1/3 of a pound is all that is necessary to exert if patient is properly positioned in the recommended lateral neutral position." other remarks: a corrective action is not required at this time as a potential root cause could not be determined based upon the information provided and without a sample.Complaint verification testing could not be performed as no sample was returned for analysis.No lot number was provided.A device history record review was performed based upon a lot number from sales history data.A device history record review was performed on the epidural catheter with no evidence to suggest a manufacturing related cause.Therefore, the potential cause of catheter kinking and being difficult to remove could not be determined based upon the information provided and without a sample.
 
Event Description
The anesthesiologist reported that the epidural catheter has formed a true knot in the tip of the catheter prior to removal, making it extremely difficult to remove.The patient's condition was reported as fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIDURAL CATHETERIZATION KIT
Type of Device
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
312 commerce place
asheboro NC 27203
Manufacturer Contact
jasmine brown
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key6210237
MDR Text Key63477527
Report Number1036844-2016-00687
Device Sequence Number1
Product Code CAZ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/13/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberSJ-05501
Device Lot NumberNO LOT GIVEN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/30/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-