MEDTRONIC NEUROSURGERY STRATA II VALVE, REGULAR SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS Back to Search Results
Catalog Number 42866
Device Problem Inaccurate Flow Rate (1249)
Patient Problem Therapeutic Response, Decreased (2271)
Event Date 12/01/2016
Event Type  Injury  
Manufacturer Narrative
The product was unavailable for return. Therefore an evaluation of the device performance was not possible. A review of the manufacturing records was not possible as no lot number was provided. All valves are 100% tested at the time of manufacture. A good faith effort will be made to obtain the applicable information relevant to the report. If information is provided in the future, a supplemental report will be issued.
 
Event Description
Iit was reported to medtronic neurosurgery that the valve was not draining like it should. According to the report, the valve was replaced and there was no injury to the patient.
 
Manufacturer Narrative
The returned valve was patent. The valve met the requirements for siphon, pressure flow, and preimplantation testing. However, the valve did not meet the requirements for reflux testing. There was proteinaceous debris observed within the interior and exterior of the valve. Debris within the valve may hold pressure-flow controlling mechanisms open resulting in fluid reflux. The instructions for use that accompany the device caution that ¿shunt obstruction may occur in any of the components of the shunt system. The system may become occluded internally due to tissue fragments, blood clots, tumor cell aggregates, bacterial colonization, or other debris. ¿ the returned valve also did not meet the requirements for leak testing due to a tear in the top of the reservoir. It is unknown how or when this damage occurred. The ifu cautions that ¿care should be taken in handling the valves as silicone has a low cut and tear resistance. ¿ the ifu also cautions that ¿improper use of instruments in the handling or implantation of shunt products may result in the cutting, slitting, crushing or breaking of components. ¿ a review of the manufacturing records was not possible as no lot number was provided. All valves are 100% tested at the time of manufacture. A good faith effort will be made to obtain the applicable information relevant to the report. If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSTRATA II VALVE, REGULAR
Type of DeviceSHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
Manufacturer (Section G)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
Manufacturer Contact
bob shokoohi
125 cremona drive
goleta, CA 93117
8055718725
MDR Report Key6211437
MDR Text Key63468315
Report Number2021898-2016-00492
Device Sequence Number1
Product Code JXG
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K042465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 12/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number42866
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/17/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received04/06/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 12/28/2016 Patient Sequence Number: 1
-
-