RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
A follow up report will be submitted once the investigation is complete.
 
Event Description
The customer reported the unit went to ventilator inoperative with error code message of data acquisition (da) pcba adc failed.The unit was in clinical at the time the reported issue was discovered; however, there was no harm to the patient or the user.
 
Manufacturer Narrative
Multiple attempts were made to follow-up with the customer to obtain patient information, however there was no response.The manufacturer's field service engineer (fse) delivered the retrofit kit, data acquisition (da) to motor controller (mc) cable/mc bracket to the customer for replacement.Multiple attempts were made to follow-up with the customer to obtain repair information, however, there was no response.If additional information is received at a later date, this complaint will be re-opened and re-evaluated accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
betty harris
3000 minuteman road
andover, MA 01810
9786871501
MDR Report Key6215524
MDR Text Key63835170
Report Number2031642-2016-03636
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Remedial Action Recall
Type of Report Initial,Followup
Report Date 12/07/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/07/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/21/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-2061-2017
Patient Sequence Number1
-
-