Model Number 980
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The service engineer (se) inspected the device and verified the reported issue.The se replaced the line interface 2 printed circuit board (pcb).The se performed extended self-testing on the device and all tests passed.
 
Event Description
It was reported that, on start up a 980 ventilator generated a serial number mismatch diagnostic message and failed the breath delivery power on self test (post).The ventilator was not in use on a patient at the time the event occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
980 VENTILATOR
Type of Device
CONTINUOUS VENTILATOR
Manufacturer (Section D)
COVIDIEN
new mervue industrial park
michael collins road
galway
EI 
Manufacturer (Section G)
COVIDIEN
new mervue industrial park
michael collins road
galway
EI  
Manufacturer Contact
ray maroofian
2101 faraday ave
carlsbad, CA 92008
7606035334
MDR Report Key6215741
MDR Text Key63823387
Report Number8020893-2016-03689
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131252
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/05/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number980
Device Catalogue Number980
Was Device Available for Evaluation? Yes
Date Manufacturer Received12/05/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/31/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-