Model Number MMT-551NAH |
Device Problem
Device Inoperable (1663)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/05/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.This mdr related to the (b)(4) manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.
|
|
Event Description
|
The customer reported via phone call of issues with their insulin pump.The customer states the pump was damaged at the reservoir compartment.The customer states it was difficult for the reservoir to remain in place.The customer's blood glucose level was unknown.The customer was advised the pump would be replaced and returned for analysis.
|
|
Manufacturer Narrative
|
The insulin pump was received with minor scratched lcd window, broken reservoir tube lip, missing reservoir tube o-ring and cracked battery tube threads.
|
|