Catalog Number 1011708-38
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Intimal Dissection (1333)
Event Date 12/12/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.The reported patient effect of dissection is listed in the xience prime stent instructions for use as a known patient effect of coronary stenting procedures.A conclusive cause for the reported patient effect, and the relationship to the product, if any, cannot be determined.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event Description
It was reported that the 2.75x38 xience prime stent was implanted in the predilated, mid left anterior descending coronary artery at 14 atmospheres.The stent was post dilated with a 2.75x18 non-compliant balloon.A distal edge dissection was observed.A 2.5x12 xience prime stent was implanted to seal the dissection.There was no adverse patient sequel.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE PRIME LL EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
CLONMEL, IRELAND REG# 9616693
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key6216555
MDR Text Key63686969
Report Number2024168-2016-09409
Device Sequence Number1
Product Code NIQ
UDI-Device Identifier08717648123580
UDI-Public(01)08717648123580(17)190720(10)6071541
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Health Professional
Type of Report Initial
Report Date 12/30/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/20/2019
Device Catalogue Number1011708-38
Device Lot Number6071541
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/26/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age48 YR
Patient Weight60
-
-