Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
Pre-op diagnosis: spinal canal stenosis procedure: posterior interbody fusion levels: l3/4/5 it was reported that intra-op, screw could not be inserted during fixation.The surgeon released the constructed implants once and placed the set screw but he could not tighten it.As a result, the surgeon performed final tightening by hand and closed the incision.Although provisional tightening was conducted without problem, when the surgeon tried to apply toque to the set screw, it was spinning.The surgeon decided not to perform breaking off the set screw and the incision was closed in provisional tightening state.The surgeon commented that screw head might have been widened, or cross thread might have occurred.The screw is remaining in the patient¿s body.No patient complications were reported as a result of the event.
|