AMEDA, INC. PURELY YOURS ULTRA; ELECTRIC BREAST PUMP Back to Search Results
Model Number 24501879
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/06/2016
Event Type  malfunction  
Manufacturer Narrative
Upon investigation, it was found that batteries were installed backwards in the battery compartment.Additional testing confirmed that batteries may leak when the batteries are placed incorrectly, with the positive and negative end reversed.
 
Event Description
Customer contacted ameda, inc.On 12/07/2016 to report an incident while using the purely yours ultra breast pump on (b)(6) 2016.She had installed 6 aa batteries inside the battery compartment of the breast pump base since she had no available electrical outlet.Customer reports black battery fluid leaking from the pump base while pumping.This has occurred in 2 separate occasions; the first event occurred 1.5 months before the recent event but was not reported at that time.In both situations, customer states no property damage, burn or injury occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PURELY YOURS ULTRA
Type of Device
ELECTRIC BREAST PUMP
Manufacturer (Section D)
AMEDA, INC.
485 half day rd.
suite 320
buffalo grove IL 60089
Manufacturer Contact
linda zager
485 half day rd.
suite 320
buffalo grove, IL 60089
8479642620
MDR Report Key6223429
MDR Text Key63883425
Report Number3009974348-2017-00240
Device Sequence Number1
Product Code HGX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973501
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Remedial Action Replace
Type of Report Initial
Report Date 12/07/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/04/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number24501879
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received12/07/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/30/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age31 YR
-
-