Device Problem
Image Display Error/Artifact (1304)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The pump has been returned to animas.Evaluation has not yet been completed.When evaluation is complete a supplemental report will be filed.No conclusion can be made at this time.
|
|
Event Description
|
On (b)(6) 2016, the reporter contacted animas, alleging that a line was present going through the pump display screen.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the issue may impact the user's ability to read some or all of the information on the screen which may result in over or under delivery.
|
|
Manufacturer Narrative
|
Follow-up #1: date of submission 02/14/2017.Device evaluation: the device has been returned and evaluated by product analysis on 01/28/2017 with the following findings: on investigation, the pump powered on with a normal display screen; there were no lines in the display screen.Investigation did not duplicate the complaint of a line through display.
|
|