Model Number CI-1500-04 |
Device Problems
Electrical /Electronic Property Problem (1198); Connection Problem (2900); Intermittent Communication Failure (4038)
|
Patient Problems
No Code Available (3191); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 06/14/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
The recipient's device is reportedly not connecting to the programming software.External equipment was exchanged, however, the issue was not resolved.
|
|
Manufacturer Narrative
|
Revision surgery will be scheduled.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The recipient has reportedly elected to not proceed with revision surgery due to financial issues.It is believed that the recipient experienced an in-situ failure.A review of the device history record was completed and no anomalies were noted.The recipient remains implanted.This is the final report.
|
|
Manufacturer Narrative
|
The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed a dented bottom cover.The photographic imaging inspection revealed a cracked hybrid and broken components.System lock could not be obtained at any spacing.The no lock condition prevented an electrical test from being performed.The device passed some of the electrical tests performed.The device passed the mechanical tests performed.The reported complaint of loss of lock with the device was verified during this analysis.The device was received with multiple cracks along the hybrid and damaged electronic components.A corrective action was implemented.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed a dented bottom cover.The photographic imaging inspection revealed a cracked hybrid and broken components.System lock could not be obtained at any spacing.The no lock condition prevented an electrical test from being performed.The device passed some of the electrical tests performed.The device passed the mechanical tests performed.The reported complaint of loss of lock with the device was verified during this analysis.The device was received with multiple cracks along the hybrid and damaged electronic components.A corrective action was implemented.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|