Device Problem Failure to Prime (1492)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The pump has not been returned to animas for evaluation. If the device is returned, an evaluation shall be completed and a supplemental report will be filed. No conclusions can be made at this time. (b)(6).
 
Event Description
On (b)(6) 2016, the reporter contacted animas alleging a prime (loss of prime) issue. It was reported that the loss of prime occurred 2 times. There is no indication that the product issue caused or contributed to an adverse event. This complaint is being reported because the issue may result in a long term cessation of insulin delivery if the user is unable to resolve the alarm.
 
Manufacturer Narrative
Follow-up # 1 date of submission 3/13/2017 device evaluation: the device has been returned and evaluated by product analysis on 2/15/2017 with the following findings: the black box data from time of the complaint has been overwritten due to continued use of the pump. Because of the continued use of the pump, the investigation was unable to confirm or duplicate the initial complaint about a loss of prime issue. The black box data from before the complaint was obtained and there was no evidence of a loss of prime issue. During testing, the pump successfully completed a rewind, load, and prime sequence and the 24 hour duration test with no errors, alarms, or warnings occurring. The force sensor calibration passed within specification. Unrelated to the initial complaint, it was observed that the battery compartment was cracked. The battery cap and cartridge cap were not returned with the pump and test caps were used to complete the investigation. Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release. (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameANIMAS VIBE
Type of DeviceINSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer (Section G)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer Contact
karin sargrad
200 lawrence dr
west chester, PA 19380-3428
4843561808
MDR Report Key6233031
MDR Text Key64587389
Report Number2531779-2017-00630
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
PMA/PMN Number
P130007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 12/19/2016
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/09/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Other Device ID Number1-MCBC-3762
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Device Age18 MO
Event Location No Information
Date Manufacturer Received12/19/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/09/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-