Model Number 6379 |
Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/16/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
The reported complaint was not confirmed.
The service repair technician (srt) set varied speed and flow settings with multiple flow sensors.
No fluctuations at any speed and flow settings were observed.
If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Manufacturer Narrative
|
(b)(4).
The field service representative (fsr) tested the device ans was unable to verify the reported issue.
The fsr did note the flow probe signal strength was only reading 36%.
The device was returned for further evaluation.
|
|
Event Description
|
It was reported that during priming of the device for a cardiopulmonary bypass procedure, the centrifugal controller had a slow flow reading, fluctuating from 2l/min - 5l/min.
As a result, an alternate device was employed.
The surgical procedure was completed successfully.
There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
During laboratory analysis, the product surveillance technician (pst) connected a centrifugal motor and flow sensor to the centrifugal control module.
Pst operated the module at flow rates from 0-9 l/min with no fluctuations in flow readings.
Operated the module for three days with no failure occurring.
If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Search Alerts/Recalls
|