• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. MERLIN PROGRAMMER; PACEMAKER PROGRAMMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. MERLIN PROGRAMMER; PACEMAKER PROGRAMMER Back to Search Results
Model Number 3650
Device Problem No Pacing (3268)
Patient Problem Dizziness (2194)
Event Date 12/12/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
While performing the ventricular capture test, the programmer failed the test and the patient was without pacing for a few seconds.The emergency vvi button needed to be pressed to pace the patient.The patient experienced dizziness during the episode.The programmer was restarted, and the test was completed.The patient was in stable condition.
 
Manufacturer Narrative
Tried several attempts with the vvi button functionality.Unable to reproduce any functionality problem of the vvi button.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN PROGRAMMER
Type of Device
PACEMAKER PROGRAMMER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6262423
MDR Text Key65512349
Report Number2017865-2017-00422
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/19/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/18/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3650
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/19/2017
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-