Device Problems
Failure to Read Input Signal (1581); Structural Problem (2506)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The pump has been returned to animas for evaluation.An evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
|
Event Description
|
On (b)(6) 2016, the reporter contacted animas, alleging a button/keypad (abb dmg prior to tactile cng) issue; audio bolus button cover was torn/damaged, then the audio bolus button became under-responsive.This complaint is being reported because the issue has the ability to result in inadvertent or incorrect boluses which may result in over or under delivery.There was no indication that the product caused or contributed to an adverse event.
|
|
Manufacturer Narrative
|
The device has been returned and evaluated by product analysis on 02/08/2017 with the following findings: during investigation, the audio bolus button cover was damaged/punctured.The complaint of the under responsive audio bolus button cover was unable to be duplicated.Unrelated to the original complaint, the pump was cracked between the case seal and the keypad cover.Additionally, the battery compartment was cracked at the case seal below the bumper.
|
|