ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Device Problems Failure to Read Input Signal (1581); Structural Problem (2506)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The pump has been returned to animas for evaluation.An evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event Description
On (b)(6) 2016, the reporter contacted animas, alleging a button/keypad (abb dmg prior to tactile cng) issue; audio bolus button cover was torn/damaged, then the audio bolus button became under-responsive.This complaint is being reported because the issue has the ability to result in inadvertent or incorrect boluses which may result in over or under delivery.There was no indication that the product caused or contributed to an adverse event.
 
Manufacturer Narrative
The device has been returned and evaluated by product analysis on 02/08/2017 with the following findings: during investigation, the audio bolus button cover was damaged/punctured.The complaint of the under responsive audio bolus button cover was unable to be duplicated.Unrelated to the original complaint, the pump was cracked between the case seal and the keypad cover.Additionally, the battery compartment was cracked at the case seal below the bumper.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Type of Device
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer (Section G)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer Contact
karin sargrad
200 lawrence dr
west chester, PA 19380-3428
4843561808
MDR Report Key6262707
MDR Text Key65509366
Report Number2531779-2017-01540
Device Sequence Number1
Product Code MDS
UDI-Device Identifier10840406100075
UDI-Public0110840406100075
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 12/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/18/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received12/22/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age24 YR
-
-