ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ALLURE QUADRA RF CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM3242
Device Problem Pacemaker Found in Back-Up Mode (1440)
Patient Problem No Patient Involvement (2645)
Event Date 03/13/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that prior to the procedure, the pulse generator was interrogated, backup vvi mode was noted.The device was used.No patient involvement.
 
Manufacturer Narrative
Final analysis confirmed the reported backup vvi operation.The root cause of the anomaly could not be determined.It was noted that the device was exposed to cold temperatures prior to its return.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURE QUADRA RF CRT-P
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6265216
MDR Text Key65571204
Report Number2017865-2017-00497
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/30/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/19/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2017
Device Model NumberPM3242
Device Lot NumberA000011309
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/05/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received03/13/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/09/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-