Device Problem
Structural Problem (2506)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device has been returned to animas.
Evaluation has not yet been completed.
When evaluation is complete a supplemental report will be filed.
No conclusion can be made at this time.
|
|
Event Description
|
On (b)(6) 2017, the reporter contacted animas, alleging a casing/condition (cracked/damaged casing) issue.
It was alleged that the battery compartment was damaged.
There was no indication that the product caused or contributed to an adverse event.
This complaint is being reported because the alleged malfunction has the ability to result in a delay in treatment or long term cessation in delivery if the damage impacts the power circuit or cartridge compartment.
|
|
Manufacturer Narrative
|
The device has been returned and evaluated by product analysis on 02/22/2017 with the following findings: during investigation, the battery compartment was cracked below the bumper pad.
The returned battery cap was able to attach to the pump.
Unrelated to the original complaint, the cartridge compartment was damaged near the threads.
The returned cartridge cap was able to attach to the pump.
|
|