• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE, INC. GSB GLOBAL SPINAL BALANCE SYSTEM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE, INC. GSB GLOBAL SPINAL BALANCE SYSTEM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Model Number 11111550
Device Problem Detachment Of Device Component (1104)
Patient Problem Failure of Implant (1924)
Event Date 03/18/2015
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned for evaluation as it remains in-situ.No revision has been completed at this time.It is unknown if the patient followed post-operative instruction.Radiographs confirmed the event.Root cause has not been determined and no conclusion can be drawn at this time; similar events have been associated with incomplete seating of components.Review of labeling notes: ".Warning cautions and precautions potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant component(s), loss of fixation, nonunion or delayed union, fracture of the vertebra." ".Care should be taken to insure that all components are ideally fixated prior to closure." ".Patient education: preoperative instructions to the patient are essential.The patient should be made aware of the limitations of the implant and potential risks of the surgery.The patient should be instructed to limit postoperative activity, as this will reduce the risk of bent, broken or loose implant components.The patient must be made aware that implant components may bend, break or loosen even though restrictions in activity are followed." ".All lock screws should be final-tightened with the counter-torque and torque t-handle.Do not final-tighten through compression instruments (e.G.C/d rack and figure 8 compressor) in the set, as the rod may not be able to normalize to the tulip.Be cautious not to over compress or distract as you can loosen the screws in the spine and potentially pull out the screw.The bulleted portion of the nose of the rod and the faceted portion of the rod (where the inserter locks down on the rod) must extend fully outside of the most inferior or most superior tulip on the construct.The set screw cannot be locked down on this unusable portion of the rod, as this may compromise the stability of the construct.All set screws should be final-tightened with the counter- torque and torque t-handle.Do not able to normalize to the tulip.Cross connectors are designed specific to the rod on constant diameter rod sections.All set screws should be final-tightened with the counter- torque and torque t-handle.Do not final-tighten through compression instruments in the set, as the rod may not be able to normalize to the tulip." ".Post-operative warnings during the postoperative phase it is of particular importance that the physician keeps the patient well informed of all procedures and treatments.Damage to the weight-bearing structures can give rise to loosening of the components, dislocation and migration as well as to other complications.To ensure the earliest possible detection of such catalysts of device dysfunction, the devices must be checked periodically postoperatively, using appropriate radiographic techniques." the device has not been returned.
 
Event Description
On (b)(6) 2015 a (b)(6) year-old female underwent a posterior spinal fusion from t10-s1.Radiographs obtained (b)(6) 2016 showed 4 sets screws appear disengaged.Revision scheduled.No patient injury has been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GSB GLOBAL SPINAL BALANCE SYSTEM
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
NUVASIVE, INC.
7475 lusk blvd
san diego CA 92121
Manufacturer (Section G)
NUVASIVE, INC.
7475 lusk blvd
san diego CA 92121
Manufacturer Contact
george panfili
7475 lusk blvd
san diego, CA 92121
8588825019
MDR Report Key6269012
MDR Text Key65465558
Report Number2031966-2016-00096
Device Sequence Number1
Product Code NKB
UDI-Device Identifier00887517227560
UDI-Public00887517227560
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/21/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model Number11111550
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/22/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age75 YR
-
-