device is a combination product.(b)(4). device evaluated by mfr: it is indicated that the device will not be returned for evaluation.A review of the batch history, historical trending, and similar complaint trending review for the product family will be conducted.If there is any further relevant information from that review, a supplemental medwatch will be filed.(b)(4).
|
Device avail.For eval, returned to mfr.On, device returned to mfr, device evaluated by mfr, eval summary attached, method codes, result codes, conclusion codes updated.Device evaluated by mfr: a synergy ii, us, mr, 2.5 x 16 mm stent delivery system (sds) proximal end including the manifold was returned for analysis, the distal section from the distal end of the port weld to the tip was not returned.The stent was deployed during the procedure.A visual and tactile examination found an extrusion break 6-7 mm distal of port site.The port weld section appeared to be slit from the beginning to the end of the port weld section returned.The bi-component bond showed no signs of damage or strain.A visual and tactile examination of the device found severe hypotube kinks along the full catheter length.The hypotube tab was pulled proximally from its original position (beginning of the port weld) for a distance of 9 mm.Dried medium resembling blood was evident in the manifold.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
|