Device Problem
Failure to Read Input Signal (1581)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The pump has been returned to animas.Evaluation has not yet been completed.When evaluation is complete a supplemental report will be filed.No conclusion can be made at this time.Initial reporter: (b)(6).
|
|
Event Description
|
On (b)(6) 2017, the reporter contacted animas, alleging a button/keypad (tactile changes/unresponsive) issue.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the issue has the ability to result in inadvertent or incorrect insulin delivery or the inability to use the pump.
|
|
Manufacturer Narrative
|
Follow-up # 1 date of submission 3/19/2017 device evaluation: the device has been returned and evaluated by product analysis on 2/23/2017 with the following findings: during visual inspection of the pump, it was observed that the keypad was intact; all the buttons were responsive during the investigation.The keypad cover was removed and there was no contamination or physical damage found.The pump was opened and there was no evidence of the moisture corrosion near keypad connector, on the display and the transceiver boards.The investigation was unable to duplicate the initial complaint about an under responsive keypad.
|
|