The pump has been returned and evaluated by product analysis on (b)(6)2016 with the following findings: evaluation revealed cgm history shows ¿no antenna¿ warning on (b)(6)2016.¿ant¿ warning is defined as pump/transmitter communication interruption.The test transmitter was paired with the pump correctly.Bg calibration of 120 mg/dl was accepted in both attempts within 2 hours.The pump and transmitter ran over night with a steady reading of 115 mg/dl within +/- 20%.No¿ant¿ icon or any eaw occurred during the investigation, unable to duplicate ¿ant¿ complaint.The pump¿s cover was removed; no intermittent condition was found to the cgm module or to the pcb.The battery compartment is cracked below the grip pad.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.(b)(6).
|