Device Problem
Structural Problem (2506)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Follow-up #1: date of submission 07/26/2017 - device evaluation: the device has been returned and evaluated by product analysis on 06/29/2017 with the following findings: the pump exhibited cracks in battery compartment from the threads to the case seal, left of grip pad, and above the grip pad to the threads.
|
|
Manufacturer Narrative
|
The device has not been returned to animas.
If the device is returned, an evaluation shall be completed and a supplemental report will be filed.
No conclusions can be made at this time.
|
|
Event Description
|
On (b)(6) 2017, the reporter contacted animas, alleging a casing/condition (cracked/damaged casing) issue.
It was alleged that the battery compartment was damaged.
There was no indication that the product caused or contributed to an adverse event.
This complaint is being reported because the alleged malfunction has the ability to result in a delay in treatment or long term cessation in delivery if the damage impacts the power circuit or cartridge compartment.
|
|