MAUDE Adverse Event Report: ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The device has been returned to animas.Evaluation has not yet been completed.When evaluation is complete a supplemental report will be filed.No conclusion can be made at this time.
 
Event Description
On (b)(6) 2017, the reporter contacted animas, alleging a power (no power) issue.There was no indication that the product caused or contributed to an adverse event.This issue is being reported because the user may be unaware that the pump has lost power, leading to under delivery.
 
Manufacturer Narrative
Follow-up #1: date of submission 03/24/2017 device evaluation: the device has been returned and evaluated by product analysis on 03/03/2017 with the following findings: a review of the black box showed power on reset events.The pump was powered on properly.The rewind, load, and prime steps were performed successfully.The pump was run for 24 hours and no power issues occurred.The power complaint could not be duplicated.Unrelated to the original complaint, the battery compartment was cracked at the threads to the left side of the grip pad down to the seal.The battery cap was able to fit for testing.Moisture was found in the battery compartment.A leak test was performed and failed due to a leak in the battery compartment.The pump was opened to find moisture on the continuous glucose monitoring module.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Type of Device
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer (Section G)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer Contact
karin sargrad
200 lawrence dr
west chester, PA 19380-3428
4843561808
MDR Report Key6278257
MDR Text Key65941720
Report Number2531779-2017-02116
Device Sequence Number1
Product Code MDS
UDI-Device Identifier10840406100051
UDI-Public0110840406100051
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 01/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/25/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/24/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received01/06/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age20 YR
-
-