Catalog Number 5531G209
Device Problems Device Operates Differently Than Expected (2913); Insufficient Information (3190)
Patient Problems Failure of Implant (1924); Damage to Ligament(s) (1952); Pain (1994); Injury (2348)
Event Date 11/01/2016
Event Type  Injury  
Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.
 
Event Description
Implanted the triathlon by tka procedure on (b)(6) 2016.Pain occurred behind knee about (b)(6) 2016.It was confirmed pcl tear and unstable insert by an arthroscope.It will be treated by a surgical procedure on (b)(6) 2017.
 
Manufacturer Narrative
The device did not return.An event regarding instability involving a triathlon insert was reported.The event was not confirmed.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as product lot details, product return, pre- and post-operative x-rays, operative reports as well as patient history and follow-up notes are needed to complete the investigation for determining a root cause.A capa trend analysis was conducted for the reported failure mode and concluded instability may result from other factors not necessarily related to the device.No further investigation is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.Not returned.
 
Event Description
Implanted the triathlon by tka procedure on (b)(6) 2016.Pain occurred behind knee about (b)(6) 2016.It was confirmed pcl tear and unstable insert by an arthroscope.It will be treated by a surgical procedure on (b)(6) 2017.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X3 TRIATHLON CS INSERT #2 9MM
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
Manufacturer Contact
keyla colon
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key6286456
MDR Text Key66053594
Report Number0002249697-2017-00353
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
PMA/PMN Number
K141056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/30/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number5531G209
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received03/09/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Weight59
-
-