Model Number 4FC12
Device Problem Air Leak (1008)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/10/2017
Event Type  malfunction  
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryo ablation procedure, while aspirating the sheath after balloon was inserted, only air was being pulled back.This occurred with three sheaths which were all replaced in addition to the balloon catheter.The procedure was continued and completed with cryo.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: device was returned and analyzed.Visual inspection of the sheath showed that the device was intact with no apparent issues.Air aspiration was reproduced when a test balloon catheter was introduced through the sheath.A dissection showed that the hemostatic valve was leaking.In conclusion, the reported issue has been confirmed through testing.The sheath failed the returned product inspection due to a leaking hemostatic valve.
 
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6296394
MDR Text Key66360341
Report Number3002648230-2017-00059
Device Sequence Number1
Product Code DRA
UDI-Device Identifier00643169688872
UDI-Public00643169688872
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 01/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/02/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/24/2018
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number42060
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/03/2017
Date Device Manufactured10/25/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-