An evaluation of the kangaroo pump was performed for the reported condition of, ¿the pump did not alarm at completion of the feeding so the patient was feed too much.¿ the unit was triaged and the customer complaint was partially verified.The unit was powered on and the post tones were heard.The unit was loaded with a feed/ flush set and allowed to run for more than 24 hours at 400ml/hr.The unit did not turn off automatically.Afterwards, the vtbd was set to deliver a specific volume.The unit indicated the feeding complete by alarming.The unit was then tested for accuracy.The unit was allowed to run for more than 15 minutes to reach steady state at 130ml/hr.The unit loaded with a new feed/ flush set and the feed rate was set to be 130ml/hr.Then, using a calibrated measuring cylinder and a stopwatch, the volume delivered by the pump was measured after 30 minutes.The volume measured was 73 ml, which is outside the specified tolerance.A trend has been identified and a corrective and preventative action has been opened to address this issue.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.If information is provided in the future, a supplemental report will be issued.
|