Catalog Number 1012449-12
Device Problems Detachment Of Device Component (1104); Kinked (1339); Difficult To Position (1467)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/09/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was received.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.
 
Event Description
It was reported that the procedure was to treat a lesion in the non-tortuous, moderately calcified, 80% stenosed mid left anterior descending (lad) coronary artery.A 3.00 x 12 mm nc trek rx balloon dilatation catheter (bdc) was selected for the procedure.During advancement of the bdc over the unspecified guide wire and through the unspecified guiding catheter with slight resistance reported with both, the shaft of the bdc kinked then separated into two pieces, mid shaft outside the anatomy.The bdc never exited the guiding catheter.The bdc was removed and was replaced with another 3.00 x 12 mm nc trek rx bdc and the procedure was successfully completed.There was no reported clinically significant delay in the procedure.There was no reported adverse patient sequela.No additional information was provided.
 
Manufacturer Narrative
(b)(4).Evaluation summary: visual, dimensional and functional inspections were performed on the returned device.The reported separation and kink were confirmed.The reported difficulty to position over the guide wire was not confirmed.The reported difficulty to position with the guiding catheter could not be replicated in a testing environment due to the condition of the returned device.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported shaft separation, kink and difficulty positioning the device with the guide wire appear to be related to circumstances of the procedure; however, the investigation was unable to determine a conclusive cause for the reported difficulty positioning with the guiding catheter.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Type of Device
CORONARY DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
EL COYOL, COSTA RICA REG# 3009031392
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key6298454
MDR Text Key66674512
Report Number2024168-2017-00902
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K110134
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 02/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/02/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2019
Device Catalogue Number1012449-12
Device Lot Number61020G1
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/16/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/09/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age56 YR
Patient Weight115
-
-