• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. CADD® ADMINISTRATION SETS; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. CADD® ADMINISTRATION SETS; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 21-7036-01
Device Problem Leak/Splash (1354)
Patient Problem Complaint, Ill-Defined (2331)
Event Date 09/07/2016
Event Type  Injury  
Manufacturer Narrative
Voluntary medwatch submission #: mw5066816.The device is currently being evaluated; smiths medical will file a follow-up report detailing the results of the evaluation once it is completed.Device evaluation in progress.
 
Event Description
It was reported that a cadd® administration set tubing was leaking where it connected to the distal end of the cassette.It was noted that the bag that the pump was in was soaked with fluid.It was unclear what the impact to the patient was.
 
Manufacturer Narrative
One cadd® 0.2 fltr spike administration set was returned for investigation.The sample was received inside a plastic bag and without its original packaging.Additionally, photographs were provided for examination; no leakage was noted during review of the photographs.Visual inspection of the returned sample was conducted at a distance of 12" to 24" and under normal conditions of illumination.During examination, a crack was found in the star tubing.Investigation determined that the observed crack was a result excessive solvent application during manufacturing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD® ADMINISTRATION SETS
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
SMITHS HEALTHCARE MFG
s.a. de c.v. ave calidad no. 4
parque industrial internaciona
tijuana, b.c. 22425
MX   22425
Manufacturer Contact
lisa perz
6000 nathan lane north
minneapolis, MN 55442
7633833074
MDR Report Key6304610
MDR Text Key66615629
Report Number3012307300-2017-00409
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K933390
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/06/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number21-7036-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/27/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received02/03/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Other; Disability;
-
-