ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) MERLIN PROGRAMMER; PACEMAKER PROGRAMMER Back to Search Results
Model Number 3650
Device Problem Improper Device Output (2953)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/19/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the device battery was found to be 2.48v with a longevity of less than 3 months when a previous interrogation about 3 months prior revealed the device battery to be 2.54v with a longevity of 1.2 years.It was noted that the previous check was an inaccurate overestimation by the programmer.The current longevity was calculated and the device had a longevity of 1-2 months, accurate to that of the current estimation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN PROGRAMMER
Type of Device
PACEMAKER PROGRAMMER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6309742
MDR Text Key66932213
Report Number2938836-2017-14088
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/07/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3650
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/19/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age52 YR
-
-