• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. PORTEX® ADULT DISPOSABLE ANESTHESIA BREATHING CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. PORTEX® ADULT DISPOSABLE ANESTHESIA BREATHING CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE) Back to Search Results
Catalog Number C37101329-NLJ
Device Problem Air Leak (1008)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/13/2017
Event Type  malfunction  
Manufacturer Narrative
Smiths medical has received the sample device.A full evaluation is anticipated, but not yet begun as the device is currently in transit to the investigation site.
 
Event Description
It was reported that a portex® adult disposable anesthesia breathing circuit had an air leak during use.The incident was observed by the user.It was noted that a pre-check was performed prior to use.No patient injury was reported.The event was considered resolved.
 
Manufacturer Narrative
One sample was returned for evaluation without its original packaging.Visual inspection of the device did not identify any tear or damage on the tube.Functional testing involved a leak test and was found to have passed.A review of the testing and inspection documents was conducted and deemed adequate.A review of the manufacturing process of a similar part was conducted.Functional testing of 32 devices from the same lot was conducted and all passed.Based on the evidence, the root cause was unable to be determined and the complaint was not confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX® ADULT DISPOSABLE ANESTHESIA BREATHING CIRCUITS
Type of Device
CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
SMITHS HEALTHCARE MFG
s.a. de c.v. ave calidad no. 4
parque industrial internaciona
tijuana, b.c. 22425
MX   22425
Manufacturer Contact
lisa perz
6000 nathan lane north
minneapolis, MN 55442
7633833074
MDR Report Key6312302
MDR Text Key66826105
Report Number3012307300-2017-00326
Device Sequence Number1
Product Code CAI
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/08/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberC37101329-NLJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received06/19/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-