Terumo cardiovascular was made aware of a perfusion article (vol.31) that addressed where the vavd vacuum may cause bubble transgression in membrane oxygenators.This was printed by sage journals.Authors include k.Nygaard, as.Thiara, c.Tronstad, ma.Ringdal, and ae.Fiane.There was no patient involvement as this was a study and not during a procedure.
|
There were no samples returned for evaluation.Review of the test method indicated in the literature found the points below to differ from the actual use of the device.-in the circuit described in the literature, there is a possibility that the negative pressure inside the reservoir may be transferred into the oxygenator module via the purge line (de-air line).During the circulation in the actual use of the oxygenator, however, the purge line must be clamped.There is, therefore, no way for the negative pressure to be transferred into the oxygenator module via the purge line.-in the circuit described in the literature, there is a possibility that the negative pressure inside the reservoir may be transferred into the oxygenator module via the arterial line of the oxygenator module.During the circulation in the actual use of the oxygenator, however, the arterial line is connected to the patient.There is, therefore, no way for the negative pressure to be transferred into the oxygenator module via the arterial line.-the literature describes the hypothesis that vavd vacuum may cause a sub atmospheric pressure in the blood component of an oxygenator when the arterial pump is stopped and this pressure difference may cause bubble transgression form the sweep gas into the blood.The ifu of this product does prohibit this action.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|